RééDOC
75 Boulevard Lobau
54042 NANCY cedex

Christelle Grandidier Documentaliste
03 83 52 67 64


F Nous contacter

0

Article

--";3! O
     

-A +A

Evaluation of methods that locate the center of the ankle for computer-assisted total knee arthroplasty

SISTON RA; DAUB AC; GIORI NJ; GOODMAN SB; DELP SL
CLIN ORTHOP RELAT RES , 2005, n° 439, p. 129-135
Doc n°: 122363
Localisation : Documentation IRR
Descripteurs : DE561 - TRAITEMENT CHIRURGICAL - GENOU, DE72 - EXPLORATION EXAMENS BILANS - CHEVILLE

Accurate alignment of the mechanical axis of the limb is important to the success of a total knee arthroplasty. Although computer-assisted navigation systems can align implants more accurately than traditional mechanical guides, the ideal technique to determine the distal end point of the mechanical axis, the center of the ankle, is unknown. In this study, we evaluated the accuracy, precision, objectivity, and speed of five anatomic methods and two kinematic methods for estimating the ankle center in 11 healthy subjects. Magnetic resonance images were used to characterize the shape of the ankle and establish the true ankle center. The most accurate and precise anatomic method was establishing the midpoint of the most medial and most lateral aspects of the malleoli (4.5 +/- 4.1 mm lateral error; 2.7 +/- 4.5 mm posterior error). A biaxial model of the ankle (2.0 +/- 6.4 mm medial error; 0.3 +/- 7.6 mm anterior error) was the most accurate kinematic method. Establishing the midpoint of the most medial and most lateral aspects of the malleoli was an accurate, precise, objective, and fast method for establishing the center of the ankle.

Langue : ANGLAIS

Mes paniers

4

Gerer mes paniers

0