A new device for in situ static and dynamic calibration of force platforms

HSIEH HJ; LU TW; CHEN SC; CHANG CM; HUNG C
GAIT POSTURE , 2011, vol. 33, n° 4, p. 701-705
Doc n°: 152702
Localisation : Documentation IRR

D.O.I. : http://dx.doi.org/DOI:10.1016/j.gaitpost.2011.03.005
Descripteurs : DF3 - ANALYSE DU MOUVEMENT

In human motion analysis, in situ calibration of the force plate is necessary to
improve the accuracy of the measured ground reaction force (GRF) and center of
pressure (COP). Few existing devices are capable of both static and dynamic
calibration of the usually non-linear GRF and COP errors, while are also easy to
move and/or set up without damaging the building. The current study developed a
small device (160 cm x 88 cm x 43 cm) with a mass of 50 kg, equipped with
auxiliary wheels and fixing suction pads for rapid deployment and easy set-up. A
PC-based controller enabled quick movement and accurate positioning of the
applied force to the calibration point. Static calibration at 100 validation
points and dynamic calibration of a force plate were performed using the device.
After correction by an artificial neural network (ANN) trained with the static
data from another 121 points, the mean errors for the GRF were all reduced from a
maximum of 0.64% to less than 0.01%, while those for the COP were all reduced
from a maximum of about 1.37 mm to less than 0.04 mm. For dynamic calibration,
the mean errors for the GRF were reduced from a maximum of 0.46% to less than
0.28%, while those for the COP were reduced from a maximum of 0.95 mm to less
than 0.11 mm. The results suggest that the calibration device with the ANN method
will be useful for obtaining more accurate GRF and COP measurements in human
motion analysis.
CI - Copyright (c) 2011 Elsevier B.V. All rights reserved.

Langue : ANGLAIS